Options and accessories
7.2 Fitting the Interface
WARNING: Switch off the instrument and
remove the AC power cord before removing
the instrument case. ANTI-STATIC
PRECAUTIONS MUST BE TAKEN WHILE
INSTALLING THE INTERFACE CARD.
1.
Remove the six screws holding the top half of the instrument case
(see figure 7.1) and lift off vertically, taking care not to stretch or
break the ground connection to the case cover (green/yellow wire).
2.
Remove the two screws holding the communications interface
blanking plate located on the rear panel.
3.
(IEEE only)
Remove the four screws holding the rear panel to allow
the interface output connector to fit.
4.
Fit the interface card in socket J18 on the instrument's main PCB,
ensure the connector is firmly located.
5.
(IEEE only)
Replace the instrument rear panel and secure the four
mounting screws.
6.
Secure the interface to the rear panel with the two screws from the
blanking plate.
7.
Refit the instrument top case and replace the six securing screws.
Figure 7.1 - Fitting an interface PCB
Page
7-2
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...