5.4.8 Selecting ext. ref. junction PRT linearization whose
temperature is measured by a PRT connected to the same channel.
1. The
[Ext]
menu option allows you to set up an external reference
junction whose temperature is measured by a PRT connected to the
same channel. When you select
[Ext]
you are prompted to choose
the PRT linearization for this PRT, the screen displays:
Standard?
IEC US/ EN-
751 JIS 751 Usr
Quit OK
2.
This lets you choose one of three standard linearizations which can
be used with uncalibrated probes. A fourth option,
[Usr],
gives you
access to the 20 probe memories for Callendar van Dusen
coefficients provided with calibrated probes.
3.
If you want to select one of the standard linearization options,
[IEC751], [US/JIS] or [EN751]
, press the appropriate function key
and then
[Quit].
Reference junction information is displayed at the
top right hand corner of the screen, e.g.
RJ Mode= Ext US/JIS
4.
If you want to access one of the probe calibrations already in
memory, press
[Usr].
The screen displays:
Enter Probe #(1-20):
Quit OK
Use the keypad to enter the probe number and then press
[Quit].
Section 5.3.5
describes how to enter user coefficients.
Operating the TTI-7-R
Page 5-13
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...