Communications Interface
Sets the trigger mode to single shot.
Selects the channel as specified by the command.
Sets the scanner mode to Off.
Sets the logging to memory mode to Off.
FETCH?
Transfer the last reading stored in the instrument's internal memory
by the INITiate command to the instrument's output buffer. After
reading, the memory is cleared and further FETCh? commands do
not return anything until a fresh INITiate command is sent.
READ?
A measurement is made and the result is sent directly to the output
buffer. The trigger mode is as set using TRIG:MODE command. If in
SINGIe mode, only one reading is taken. If in INFinite mode, readings
continue, with all values sent to the output buffer.