7.6.3 MEASure Command
These work similar to the CON Figure commands, except that an
immediate measurement is performed with the result being sent to the
output buffer.
MEASure:CHANnel? {channel}
Selects the channel to be measured (use CONF:TEMP to configure the
measurement parameters).
{channel}
A0, A1, A2, A3, A4
B0, B1, B2, B3, B4
Ch1 -Ch2
Terminates the current measurement cycle and scanning routine.
Sets the trigger mode to single shot.
Selects the channel as specified by the command.
Sets the scanner mode to Off.
Sets the logging to memory mode to Off.
Initiates a single measurement and fetches a single reading.
MEASure:TEMPerature:TC? {type},{rj mode},{ext rj standard}
Sets the selected channel for a thermocouple measurement
according to the following parameters.
{type}
B,C,D,E,J,K,N,R,S,
T
{rj mode}
Off, Int, Ext
For rj mode set to Off or Int
{ext rj standard}
0
For rj mode set to Ext
{ext rj standard}
1
(IEC751)
2
(US/JIS)
3
(EN60751)
4-23
(User Probe 1-20)
Note: For {rj mode} set to Off or Int, the {ext rj standard}
parameter must be set to 0 otherwise the command is not
recognised.
Options and accessories
Page
7-18
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...