Options and accessories
CONFigure:TEMPerature:TC {type},{rj mode},{ext rj standard}
Sets the selected channel for a thermocouple measurement according to
the following parameters.
{type}
B,C,D,E,J,K,N,R,S,
T
{rj mode}
Off, Int, Ext
For rj mode set to Off or
Nt
{ext rj standard}
0
For rj mode set to Ext
{ext rj standard}
1
(IEC751)
2
(US/JIS)
3
(EN60751)
4-23
(User Probe 1-20)
Note: For {rj mode} set to Off or Int, the {ext rj standard} parameter
must be set to 0 otherwise the command is not recognised.
Terminates the current measurement cycle and scanning routine.
Sets the trigger mode to single shot.
Selects the channel as specified by the command.
Sets the scanner mode to Off.
Sets the logging to memory mode to Off.
CONFigure:TEMPerature:RTD{type},{standard},{con},{current
mode},{root 2}
Sets the selected channel for an RTD measurement according to the
following parameters.
{type}
Pt100
{standard}
1
(IEC751)
2
(US/JIS)
3
(EN60751)
4-23
(User Probe 1-20)
{con}
3, 4
(3 or 4 wire connectivity)
{current mode}
I, -I, AVE
{root 2}
0, 1
0 = off, 1 = on
Terminates the current measurement cycle and scanning routine.
Page
7-16
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...