Options and accessories
7.15 Analogue Output
The TTI-7-R analogue output provides an output voltage in the range
of ±2 Volts that is directly proportional to the displayed reading and
selected scaling factor. The analogue output is updated on
completion of every measurement cycle and is obtained from the
digital measurement by digital to analogue conversion.
7.15.1 Specification
Resolution 1°C
Sensitivity 1 mV/
°C
Analogue Output range ±2V F.S.
7.15.2 Default Settings
On power up, the TTI-7-R will initialise with the following settings:
Measurement units
deg C
Measurement PRT Input
PRT Input A
Resolution Low
resolution
7.15.3 Analogue Output Connection
Connection to the analogue output is via an insulated BNC output
socket on the rear panel of the TTI-7-R . Connection to peripheral
equipment should be made via a suitable twin core, twisted pair
screened cable to the instrumentation input amplifier of the
peripheral equipment, (see Figure 7.1).
Figure 7.1 - Twin Core, Twisted Pair, Screened Cable
Page 7-42
Summary of Contents for TTI-7-R
Page 16: ...Setting up the TTI 7 R This page has been left blank intentionally Page 2 4...
Page 34: ...Measuring Temperature This page has been left blank intentionally Page 4 8...
Page 66: ...This page has been left blank intentionally Page 5 32 Operating the TTI 7 R...
Page 68: ...Options and accessories This page has been left blank intentionally Page 6 2...
Page 116: ...Specifications This page has been left blank intentionally Page 9 4...
Page 118: ...Communications Interface This page has been left blank intentionally Page 10 2...
Page 120: ...This page has been left blank intentionally Page 11 2...