3-6-4- End of loading
•
When all the samples have been installed, close the drawer.
Removable trays
If the analyser is initialised but not in the run cycle, the compartment remains unlocked and the assays will be
carried out as soon as the run cycle is started. Each on-board tray will be assayed whether defined for
automatic start or not.
If the analyser is in the run cycle, the compartment locks automatically.
For the trays defined for automatic start (in
Rack Configuration
menu), any new assays waiting to be carried
out are added to the analyser’s workload.
For the others, the new assays will be added to the analyser’s workload after clicking on the corresponding
button.
Fixed tray
If the analyser is initialised but not in the run cycle, the drawer remains unlocked and the assays will be carried
out as soon as the run cycle is started.
If the analyser is in the run cycle, the drawer locks automatically and any new assays waiting to be carried out
are added to the analyser’s workload.
Use
3-6- L
oading of Calibrators, Controls or Samples in the Sample Compartment (continued)
IDS-iSYS User Manual - Revision M1
Use 3-6
66
Software version V 14