3-3-2- Installation of external ancillary reagents
Description
Colour code
Field
IDS-iSYS System Liquid
(5 litres)
White.
All.
IDS-iSYS Wash (10 litres)
Blue.
Immunoassay.
IDS-iSYS D-Sorb (1 litre)
Black.
All.
AP Substrate
Green
Immunoassay*.
*Depending on the analyser configuration
Installation of IDS-iSYS System Liquid, Wash S and D-Sorb
•
Remove the lid and level sensor from the container to be replaced.
•
Remove the adhesive barcode label for SYST L and WASH container.
•
Scan the new ancillary reagent barcode label with the reader located on the front of the machine.
The analyser will beep for the first time.
•
Within 10 seconds, put the level sensor in position within the new ancillary reagent for the IDS-iSYS D-Sorb
and IDS-iSYS Wash solutions.
The analyser will beep a second time.
Replace the lid and the plunger of the IDS-iSYS System Liquid.
•
Priming of the IDS-iSYS System Liquid takes place automatically.
•
For other ancillaries, a message proposing an automatic priming is displayed:
•
Click on
to confirm the priming of the product(s).
The analyser starts the priming sequence.
Comment: The automatic priming can be cancelled by clicking on
. In this case, the priming will
take place automatically when starting a run cycle.
•
If the ancillary barcode is not read by the barcode reader, click on the ancillary reagent then use the
keyboard to enter the identifier.
!
Refer to the catalogue for the product references of the ancillary reagents to be ordered.
Use
3-3- Installation of Ancillary Reagents (continued)
IDS-iSYS User Manual - Revision M1
Use 3-3
48
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...