With all modules fitted the analyser comprises the following key components:
7
2
4
1
5
8
3
6
1.
A refrigerated reagent compartment, consisting of 15 rails containing the Immunoassay or Biochemistry
racks as well as a specialised rack for chronometric reagents.
2.
A compartment for samples , calibrators and controls with, depending on the analyser configuration, a
fixed sample tray with 64 positions , or a rack containing 6 removable trays each with 20 positions.
3.
A pipetting arm that pipettes both reagents and samples.
4.
A thermo-regulated carousel set at 37°C with 90 positions for disposable cuvettes. Incorporated into the
carousel is the spectrophotometer, used for measuring absorbancy from certain Immunoassay,
Biochemistry and Turbidimetry reactions.
5.
A sedimentation module for magnetic particles.
6.
An automatic cuvette loader, holding 960 cuvettes at a time (pre-formed as a cube).
7.
A luminometer measuring luminescence in Immunoassay reactions (Immunoassay version).
8.
Four washers for washing magnetic particles.
The barcode reader integrated into the reagent compartment identifies reagent cartridges supplied by IDS.
A second barcode reader, located on the front of the analyser identifies ancillary reagents.
Samples, calibrators and controls are identified or by the reader located on the front for analysers with a fixed
sample tray, either by a reader integrated into the sample compartment for analysers with removable sample
trays.
Operating principle
1-2- General Overview (continued)
IDS-iSYS User Manual - Revision M1
Operating principle 1-2
4
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...