The probe will be placed over the reference position, located to the left of the rinsing well.
The probe is moved using the arrow buttons on the screen :
•
Adjust first the XY positions of the probe, using the corresponding movement arrows (
Left
,
Right
,
Forward
,
Backward
). The probe must be centered on the pin of the target.
•
Then adjust the Z position: click on the
Down (1 mm)
button twice.
•
Adjust the probe position using repeatedly the
Down (0,1 mm)
button. The probe must be adjusted just
above the center pin of the target.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Gently touch the probe: the probe must slightly resist to the lateral movement.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
!
The probe can be moved using
Up
and
Down
buttons when the lid is open.
The
Left
,
Right
,
Forward
,
Backward
buttons can only move once to each direction when the
lid is open.
•
When the adjustment is completed, remove the lid support tool and close the lid (see page 36).
•
Click on
.
Maintenance
5-4-
Analyser Interventions
(continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-4
134
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...