5-4-13- Repeat an automatic maintenance
Certain maintenance activities are carried out automatically by the system, without user’s intervention.
If a fault prevents the automatic maintenance being carried out in full, an error message is displayed. A
Warning message remains in the
WARNING
window until a new maintenance is complete without any error.
When this maintenance cannot be carried out automatically by the system, a new maintenance can be
requested by the user.
5-4-13-1- Washer needle cleaning
This automatic maintenance takes place when the system is placed in standby mode.
After cleaning the sampling probe using D-SORB solution, four cuvettes are prepared, each filled with 500 µl of
this solution (injected in two steps). The cuvettes are then transferred to the four washers: the aspiration needle
of the washer is cleaned when aspirating the D-SORB solution. This step is followed by a washing step. This
maintenance takes around 3 minutes et uses 3 ml de solution D-SORB.
•
Check that D-SORB and WASH S solutions are present on the system, both with su
ffi
cient volumes..
•
From the MAINTENANCE menu, select
REPEAT AN AUTOMATIC MAINTENANCE,
then select:
As soon as the menu option is selected, the maintenance is carried out by the system.
•
When maintenance is completed without any error, the date and time are recorded in the journal of events (in
the SYSTEM RECORDS tab).
5-4-13-2- Cleaning the probe
This automatic maintenance takes place when the system is placed in standby mode.
Certain assays performed on the system require a daily cleaning of the sampling probe with the
Immunocleaner solution. This product is supplied in the form of a reagent cartridge and must be installed in
the reagent compartment.
During this maintenance, the probe performs three times the following cycle : pipetting 300 µl of
Immunocleaner solution, waiting for 30 seconds above the rinsing well, rinsing. The final rinsing uses D-Sorb
solution.
•
Check that IDS-iSYS D-Sorb and Immunocleaner solutions are present in su
ffi
cient quantities on the
analyser.
•
From the MAINTENANCE menu, select
REPEAT AN AUTOMATIC MAINTENANCE,
then select:
As soon as the menu option is selected, the maintenance is carried out by the system.
Maintenance
5-4-
Analyser Interventions
(continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-4
145
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...