Open the lid and insert the lid support tool (see pages 35 and 36). The sampling arm stays at its position.
•
Install the specific protective tray under the probe while holding the sampling head.
•
Remove the clot under the probe.
•
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Remove the protective tray.
•
Remove the lid support tool and close the lid (see page 36), then click on
. The cycle
resumes at the end of the rinse.
Probe blockage alarm
If the probe becomes clogged when dispensing into a cuvette or when rinsing, the arm stops at the position
where the fault is detected.
•
Click on
.
A message confirming the request displays:
•
To stop the cycle, click on
.
If assays are under way, the analyser finishes these before stopping.
•
Then, select the fault and click on
.
The following message displays:
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Install the specific protective tray under the probe while holding the sampling head.
•
Unscrew the probe.
•
Unclog the probe (see Section 5,
5-4-11- Unclogging the sample probe
, page 141).
•
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Click on
. The following message displays:
Use
3-8-
Assays
(continued)
IDS-iSYS User Manual - Revision M1
Use 3-8
74
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...