Message
Meaning
CV tolerance has failed on
at least one of the
calibrators
Immunoassay
The CV calculated for one of the calibrators replicates is higher than the
percentage specified in the analytical configuration.
The
CVM
message is displayed with the corresponding calibrator in the
tab.
Control out of range
All fields
The control carried out with the calibration has not been calculated or is
outside the limits of acceptability.
Determination coe
ffi
cient
too low for one assay
Biochemistry
The slope of the kinetics calculated for one of the calibrators was calculated
on a non-linear section of the curve.
No control for checking
All fields
The control measurements requested with the calibration have not been
performed. Calibration is not considered to be valid.
Out of the activity limits
Biochemistry
The slope of the kinetics calculated for one of the calibrators is higher than
the limit value of activity specified in the analytical configuration, or enzymatic
hyperactivity was detected.
Messages
4-1-
Messages Associated with Calibrations (continued)
IDS-iSYS User Manual - Revision M1
Messages 4-1
106
Software version V 14