5-4- Analyser Interventions
5-4-1- Replacement of lamp
(only for systems using a spectrophotometer with an halogen lamp)
WARNING:
RISK OF BURNS
Do not handle the lamp to be replaced immediately after turning o
ff
the analyser.
Before handling, allow the lamp to cool for approximately 10 minutes after switching
the analyser o
ff
.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Switch the analyser o
ff
and unplug the power cable.
•
Remove the bottom partition at the rear of the rinsing well.
•
Disconnect the lamp from its supply.
•
Unscrew the securing screws located on both sides of the lamp in order to remove the lamp-support unit.
•
Remove the old lamp.
WARNING:
RISK OF BURNS
Do not touch the lamp with your fingers.
•
Wipe the new lamp with a soft cloth.
!
Do not touch the lamp with your fingers.
•
Put the new lamp and its base in place, with the guide pin positioned downwards.
•
Tighten the securing screws using a screwdriver.
•
Connect the lamp to its power supply.
•
Replace the partition.
•
Remove the lid support tool and close the lid (see page 36).
•
Switch the analyser back on and start it up.
Maintenance
5-4-
Analyser Interventions
IDS-iSYS User Manual - Revision M1
Maintenance 5-4
126
Software version V 14