5-1-1- General Maintenance
5-1-1-1- Checking reagent drawer and Plexiglas
®
!
Do not use any spray products on the analyser
.
•
Open the reagent compartment drawer.
•
Check that no trace of reagent is present on the base plate of the compartment If traces of reagent are
evident, use absorbent paper to clean the rails.
•
Then use a disinfectant wipe/solution suitable for medical devices.
•
Check that there are no traces of liquid on the Plexiglas
®
over the reagent compartment. If traces of liquid
are evident, use absorbent paper to remove them.
•
Close the drawer.
•
When maintenance is complete, click on the corresponding
, then confirm the
maintenance by clicking on YES.
5-1-1-2- Checking sample drawer and Plexiglas
®
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
These parts of the analyser are in contact with biological samples. There is, therefore,
a potential risk of infection.
Wear disposable gloves for all handling procedures.
!
Do not use any spray products on the analyser
.
•
Open the sample compartment drawer.
•
Check that no trace of liquid is present on the surface of the fixed or removable sample tray. If traces of
liquid are evident, use absorbent paper to clean the surface of the sample tray. Then use a disinfectant wipe/
solution suitable for medical devices.
•
Close the drawer.
•
Open the lid (see page 35) and check that there are no traces of liquid on the Plexiglas
®
over the sample
tray. If traces of liquid are evident, use absorbent paper to remove.
•
Remove the lid support tool and close the lid (see page 36).
•
When maintenance is complete, click on the corresponding
, then confirm the
maintenance by clicking on YES.
Maintenance
5-1- Daily Maintenance (continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-1
116
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...