ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Specimen Descriptor
S
This field consists of two sections:
•
The sample type which can take for the IDS-
iSYS the following values:
•
U = Urine.
•
S = Serum.
•
O = Other.
•
The sample source (place where the sample
was taken) is not used.
Ordering Physician
Doctor
Physician Telephone
Number
Not used.
User Field 1
Not used.
User Field 2
Not used.
Laboratory Field 1
Not used.
Laboratory Field 2
Not used.
Date and Time Result
Reported or Last Modified
Not used.
Instrument Charge to
Computer System
Not used.
Instrument Section ID
Not used.
Report Type
O
This field may take the values:
•
O
= Order.
•
C = Correction.
•
P = Preliminary results.
•
F = Final results.
•
X = Cancel.
•
I = Instrument pending (in response to a
request).
•
Y = No order on record for this test
(in response to a request).
•
Z
= No record for this patient
(in response to a request).
•
Q
= Response to a request.
In the device, C is managed when a clot is
detected on a sample (see page 67).
Reserved Field
Not used.
Location or Ward of
Specimen Collection
Not used.
Nosocomial Infection Flag
Not used.
Specimen Service
Not used.
Specimen Institution
Not used.
4- ASTM Compatible V3 Version
4-4- The Order Block (continued)
IDS-iSYS Connection Protocol - Revision N1
4- ASTM Compatible V3 Version
45
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...