3-10-2- Result of controls
•
Click on the position occupied by the control.
•
The results obtained for this control are displayed along with the programmed limits of acceptance.
•
A result outside the limits is displayed in red.
•
A result which is not calculated is replaced by
* * * * *
.
•
The results of the controls are automatically stored and recorded in Levey-Jennings charts and inspected
using the Westgard rules, if selected.
•
The results of the control can be printed out from its position.
•
If a control is outside the acceptable limits, the status of the calibration is displayed as
Invalid.
•
In this case, request a new calibration.
Westgard Rules Inspection
•
If WESTGARD rules are selected for the test, the rules defined in the analytical configuration will be tested in
the following order:
Control
data
12s
13s
In control -> Accept run
R 4s
41s
10x
Out of control -> reject run
Yes
No
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
22s
The violation of certain rules will generate automatic requests. In this event, the
icon flashes in the
WARNING
button.
•
The violation of 2
2S
, 4
1S
, 10
X
rules will automatically generate a request of calibration.
•
Click on the
WARNING
button to open the window.
•
Click on
to request the calibration and the controls. If necessary, install the required calibrators and
controls on the analyser.
•
To cancel the calibration request, click on
. This action will be detailed in the Records.
•
Then release to restart the test.
•
All the results are identified by the message
WE!.
•
If another rule is violated, the test is stopped.
•
Perform the necessary actions to correct the problem and request a calibration or a single control.
•
Each time a violation occurs, a message displayed in the WARNING window allows the release of the
violation, whatever rule is violated.
•
To release a test, click
. This action will be detailed in the Records. All the results are identified by the
message
WE!.
Use
3-10-
Results (continued)
IDS-iSYS User Manual - Revision M1
Use 3-10
85
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...