Function application buttons
Starts the run cycle.
Accesses calibration and quality control requests.
Displays the current calibration for each on-board lot of reagent.
Provides access to the work list and generated results. Contains
profiles sent by the Laboratory Information System (LIS).
The work list contains the details of the profiles completed or in
progress.
When the completed profile has been stored, the results are displayed
in the work list until the associated sample is removed. The results are
automatically stored provided the tests complete without any errors/
faults.
Allows manual entry of a barcode identifier for samples or reagents in
the event of a misreading by the integrated barcode readers.
Allows access to information regarding the ancillary reagents
connected to the system, and the status of the solid and liquid
waste.
The level of each element is displayed. The following detailed
information can be displayed by clicking on the reagent button:
•
Identifier and name of the reagent.
•
Lot number.
•
Number of the container in the lot.
•
Type of container.
•
Lot expiry date.
•
In-use stability*.
•
Cumulative time on-board*.
•
On-board expiry date*.
•
Status of the reagent.
•
Available volume (or number of cuvettes).
This menu allows the user to manually enter a barcode identifier in
the event of misreading by the integrated barcode reader.
Clicking on the liquid or solid wastes allows the user to confirm
emptying.
* not managed for cuvettes
Unlock lid button
(only for analyser equipped with a lid locking system)
This button is displayed only when the analyser is equipped with a lid
locking system.
In this case, the lid is continuously locked. As the lid must be opened
for maintenance tasks or interventions, access to the lid is authorised
when the analyser is not in cycle mode.
The button, when displayed, a
ll
ows the user to unlock the lid before
opening.
User Interface Software
2-2-
Main Screen (continued)
IDS-iSYS User Manual - Revision M1
User Interface Software 2-2
22
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...