6-1- Resolving Errors in Cartridge Check System (CCS)
Problem
Possible cause & corrective action
CCS1 %CV outside limits
1 - Presence of air bubbles in the sampling fluidic circuit. Check for air bubbles
in the tubing circuit. Carry out a full priming of the arm.
2 - Check for the presence of bubbles in the reagents: there should be none.
3 - Check that the probe is not blocked or bent. Carry out a decontamination
of the probe (see Section 5-1-1-4, page 117), if applicable. If the probe is bent,
replace it and adjust the XYZ reference position (see Section 5-4-2, page 127).
4 - Check for air bubbles on the IDS-iSYS Triggers A and B tubing circuits.
Check for air bubbles in the tubing circuits.
5 - Repeat the CCS1 test.
6 - If the problem persists, contact IDS Service & Support personnel.
CCS2 %CV outside limits
1 - Presence of air bubbles in the sampling fluidic circuit. Check for air bubbles
in the tubing circuit. Carry out a full priming of the arm.
2 - Check for the presence of bubbles in the reagents: there should be none.
3 - Check that the probe is not blocked or bent. Carry out a decontamination
of the probe (see Section 5-1-1-4, page 117), if applicable. If the probe is bent,
replace it and adjust the XYZ reference position (see Section 5-4-2, page 127).
4 - Check for air bubbles in the IDS-iSYS Triggers A and B tubing.
5 - Repeat the CCS2 test.
6 - If the problem persists, contact IDS Service & Support personnel.
CCS3x Value outside limits
1 - Presence of residual IDS-iSYS Wash for the associated washer: carry out a
full priming of all washers.
2 - Check for air bubbles in the IDS-iSYS Triggers A and B tubing.
3 - Check IDS-iSYS Cuvettes and IDS-iSYS Triggers by carrying out a
CCS4 test.
4 - If CCS4 is within the limits, replace the IDS-iSYS Wash Solution in use.
Carry out full priming of washers and repeat CCS3 tests.
5- If CCS4 exceeds the limits, replace the IDS-iSYS Cuvettes and IDS-iSYS
Triggers in use.
5 - If the problem persists, contact IDS Service & Support personnel.
Problems & Corrective Action
6-1-
Resolving Errors in Cartridge Check System (CCS)
IDS-iSYS User Manual - Revision M1
Problems & Corrective Action 6-1
149
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...