ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Result abnormal Flags
N
This field may take the values:
•
L
= Below low normal.
•
H
= Above high normal.
•
LL = Below alarm low.
•
HH = Above alarm high.
•
<
of absolute low (Field of measurement).
•
>
of absolute high (Field of measurement).
•
N
= Normal.
•
A
= Abnormal.
•
U = Significant change Up.
•
D = Significant change Down.
•
B = Better.
•
W = Worse.
L, H, <, >, A and N are managed by IDS-iSYS.
All associated messages are sent in a Comment
block.
Nature of Abnormality Testing
S
This field forces S or A for the IDS-iSYS can take
the values:
•
A
= Age.
•
S
= Sex.
•
R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
Result Status
F
This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
•
C = Correction.
•
P
= Preliminary results.
•
F
= Final results.
•
X = Request not honored.
•
I = Results pending.
•
S = Partial results.
Date of change in Instrument
Normative Values
Not used.
Operator ID
Not used.
Date and Time Test Started
Not used.
Date and Time Test
Completed
20111123171733
Date and time of result of test on the GUI.
Instrument Identification
IDS-iSYS B300B0378
Name of Serial number.
3- ASTM Compatible V2 Version
3-5- The Result Block (continued)
IDS-iSYS Connection Protocol - Revision N1
3- ASTM Compatible V2 Version
28
Software version V 14
Summary of Contents for IDS-iSYS
Page 7: ......
Page 13: ...SECTION 1 Operating principle...
Page 30: ...SECTION 2 User Interface Software...
Page 48: ...SECTION 3 Use...
Page 119: ...SECTION 4 Messages...
Page 130: ...SECTION 5 Maintenance...
Page 165: ...SECTION 6 Problems Corrective Action...
Page 169: ...SECTION 7 System Configuration...
Page 182: ...Appendices...