Installation of AP Substrate
•
Remove the cap from the new ancillary reagent bottle.
•
If necessary, take out the bung and remove the empty bottle.
•
Scan the ancillary with the barcode reader.
The analyser will beep for the first time.
•
Within 10 seconds, install the new bottle in its position.
The analyser will beep a second time.
•
Install the supply line into the bottle.
•
If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the identifier.
•
Once a new bottle is installed, a message is displayed:
•
Click on
to confirm the priming of the AP substrate.
The analyser starts the AP Substrate priming sequence.
Comment: The automatic priming of the AP Substrate circuit can be cancelled by clicking on
. In
this case, the priming will take place automatically when starting a run cycle.
Use
3-3- Installation of Ancillary Reagents (continued)
IDS-iSYS User Manual - Revision M1
Use 3-3
49
Software version V 14