Determining System Measurement Uncertainties
In any measurement, certain measurement errors associated with the system add
uncertainty
to the measured results This uncertainty
how accurately a device under test
can
be measured.
Network analysis measurement errors can be separated
two types: raw and residual.
The raw error terms are the errors associated
the uncorrected system that are called
systematic (repeatable), random (non-repeatable), and drift errors. The residual error terms are
the errors that remain after a measurement calibration.
The error correction procedure,
called measurement calibration, measures a set of
calibration devices
known characteristics. It uses the measurement results to effectively
remove systematic errors, using the vector math capabilities of the analyzer. The residual
systematic errors remain after error correction, primarily due to the limitations of how
accurately the electrical characteristics of the calibration devices can be defined and
determined. Also, the random (non-repeatable) and drift errors, cannot be corrected because
they cannot be quantified and measured during the measurement calibration and device
measurement. However, the effects of random errors can be reduced through averaging. The
averaging may then be reduced for device measurement. The residual systematic errors along
with the random and drift errors continue to affect measurements after error correction,
adding an uncertainty to the measurement results Therefore, measurement uncertainty is
defined as the combination of the residual systematic (repeatable), random (non-repeatable),
and drift errors in the measurement system after error correction.
The following measurement uncertainty equations and system error models (flowgraphs) show
the relationship of the systematic, random, and drift errors. These are useful for predicting
measurement performance.
Sources of Measurement Errors
The residual (after measurement calibration) systematic errors result from imperfections
in the calibration standards, the connector interface, the interconnecting cables, and the
instrumentation. All measurements are affected by dynamic accuracy and frequency error
effects For reflection measurements, the associated residual errors are effective directivity,
effective source match, and effective reflection tracking. For transmission measurements,
the additional residual errors are effective crosstalk, effective load match, and effective
transmission tracking.
The listing below shows the abbreviations used for residual systematic errors that are in the
error models and uncertainty equations.
n
Efd, Erd = effective directivity
n
Efs, Ers effective source match
n
Efr, Err effective reflection tracking
Determining System Measurement Uncertainties
Summary of Contents for HP 8753E
Page 61: ......
Page 150: ......
Page 187: ......
Page 197: ...4 l 8 Printing Plotting and Saving Measurement Results ...
Page 245: ......
Page 249: ......
Page 256: ......
Page 277: ...4 lb enter the number of sweeps press Optimizing Measurement Results 5 55 ...
Page 399: ......
Page 475: ......
Page 476: ...pb611ld 8 2 Menu Maps ...
Page 477: ......
Page 478: ...84 Menu Maps ...
Page 480: ......
Page 487: ......
Page 488: ...8 14 MenuMaps ...
Page 489: ...I r RETURN pg657e Menu Maps 8 15 ...
Page 565: ... Ittble 9 2 Softkey Locations Softkey Front Panel Access Key B 76 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 77 ...
Page 567: ...9 79 Key Definitions ...
Page 568: ...Key Definitions 8 78 ...
Page 569: ... Ihble 9 2 Softkey Locations continued Softkey A i _ 8 80 Key Definitions ...
Page 570: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 81 ...
Page 572: ...lhble 9 2 Softkey Lacations continued Softkey Front Panel Access Key Key Definitions 9 83 ...
Page 573: ... I ble 9 2 Softkey Locations continued Softkey Front Panel Access K e y 84 I ey Definitions ...
Page 574: ... lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions g 86 ...
Page 575: ... lhble 9 2 Softkey Locations continued Softkey FFont Panel Access Key 9 86 Kay Definitions ...
Page 576: ... ble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 87 ...
Page 577: ... lhble 9 2 Softkey Locations continued Softkey Front Fanel Access Key B 88 Key Definitions ...
Page 578: ...l hble 9 2 Softkey Locations continued Softkey Key Definitiins 8 W ...
Page 579: ... lhble 9 2 Softkey Locations continued Softkey Front Rmel Access Key WI Key Definitions ...
Page 580: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions Ml ...
Page 581: ... able 9 2 Softkey Locations continued Key Definitions ...
Page 582: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 943 ...
Page 583: ...7hble 9 2 Softkey Locations continued I Softkey Front Panel Access Key 944 Key Definitions ...
Page 584: ... Ihble 9 2 Softkey Locations continued Softkey Key Definitions 9 96 ...
Page 671: ......
Page 672: ......
Page 673: ......