Figure 6-54. Typical Measurement Set up
If the device measurement requires bias, it will be necessary to add external bias tees between
the
attenuators and the fixture. The internal bias tees of the analyzer will not pass the
bias properly through the external
attenuators, Be sure to calibrate with the external bias
tees in place (no bias applied during calibration) to remove their effect from the measurement.
Because the bias tees must be placed after the attenuators, they essentially become part of the
fixture. Their mismatch effects are the same for source match and load match, so the TRL CAL
routine will correct for their effects Although the fixed attenuators improve the raw mismatch
of the network analyzer system, they also degrade the
measurement dynamic range.
This effective mismatch of the system after calibration has the biggest effect on reflection
measurements of highly reflective devices, Likewise, for well matched devices, the effects of
mismatch are negligible. This can be shown by the following approximation:
Reflection magnitude uncertainty
+
+
+
Transmission magnitude uncertainty = +
+
+
where:
En = effective directivity
En = effective reflection tracking
Es = effective source match
= effective load match
effective crosstalk
= effective transmission tracking
= S-parameters of the device under test
Application and
Concepts
Summary of Contents for HP 8753E
Page 61: ......
Page 150: ......
Page 187: ......
Page 197: ...4 l 8 Printing Plotting and Saving Measurement Results ...
Page 245: ......
Page 249: ......
Page 256: ......
Page 277: ...4 lb enter the number of sweeps press Optimizing Measurement Results 5 55 ...
Page 399: ......
Page 475: ......
Page 476: ...pb611ld 8 2 Menu Maps ...
Page 477: ......
Page 478: ...84 Menu Maps ...
Page 480: ......
Page 487: ......
Page 488: ...8 14 MenuMaps ...
Page 489: ...I r RETURN pg657e Menu Maps 8 15 ...
Page 565: ... Ittble 9 2 Softkey Locations Softkey Front Panel Access Key B 76 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 77 ...
Page 567: ...9 79 Key Definitions ...
Page 568: ...Key Definitions 8 78 ...
Page 569: ... Ihble 9 2 Softkey Locations continued Softkey A i _ 8 80 Key Definitions ...
Page 570: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 81 ...
Page 572: ...lhble 9 2 Softkey Lacations continued Softkey Front Panel Access Key Key Definitions 9 83 ...
Page 573: ... I ble 9 2 Softkey Locations continued Softkey Front Panel Access K e y 84 I ey Definitions ...
Page 574: ... lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions g 86 ...
Page 575: ... lhble 9 2 Softkey Locations continued Softkey FFont Panel Access Key 9 86 Kay Definitions ...
Page 576: ... ble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 87 ...
Page 577: ... lhble 9 2 Softkey Locations continued Softkey Front Fanel Access Key B 88 Key Definitions ...
Page 578: ...l hble 9 2 Softkey Locations continued Softkey Key Definitiins 8 W ...
Page 579: ... lhble 9 2 Softkey Locations continued Softkey Front Rmel Access Key WI Key Definitions ...
Page 580: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions Ml ...
Page 581: ... able 9 2 Softkey Locations continued Key Definitions ...
Page 582: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 943 ...
Page 583: ...7hble 9 2 Softkey Locations continued I Softkey Front Panel Access Key 944 Key Definitions ...
Page 584: ... Ihble 9 2 Softkey Locations continued Softkey Key Definitions 9 96 ...
Page 671: ......
Page 672: ......
Page 673: ......