Measurement Calibration
Measurement calibration is an accuracy enhancement procedure that effectively removes the
system errors that cause uncertainty in measuring a test device. It measures known standard
devices, and uses the results of these measurements to characterize the system.
This section discusses the following topics:
of accuracy enhancement
n
causes of measurement errors
characterization of microwave systematic errors
n
calibration considerations
n
effectiveness of accuracy enhancement
n
correcting for measurement errors
n
ensuring a valid calibration
n
modifying calibration kits
calibration
n
power meter calibration
calibrating for noninsertable devices
What Is Accuracy Enhancement?
A perfect measurement system would have
dynamic range, isolation, and directivity
characteristics, no impedance mismatches in any part of the test setup, and flat frequency
response. In any high frequency measurement there are measurement errors associated with
the system that contribute uncertainty to the results. Parts of the measurement setup such as
interconnecting cables and signal-separation devices (as well as the analyzer itself) introduce
variations in magnitude and phase that can mask the actual performance of the test device.
Vector accuracy enhancement, also known as measurement calibration or error-correction,
provides the means to simulate a nearly perfect measurement system.
For example, crosstalk due to the channel isolation characteristics of the analyzer can
contribute an error equal to the transmission signal of a high-loss test device. For reflection
measurements, the primary limitation of dynamic range is the directivity of the test setup.
The measurement system cannot distinguish the true value of the signal reflected by the test
device from the signal arriving at the receiver input due to leakage in the system. For both
transmission and reflection measurements, impedance mismatches within the test setup cause
measurement uncertainties that appear as ripples superimposed on the measured data.
Error-correction simulates an improved analyzer system. During the measurement calibration
process, the analyzer measures the magnitude and phase responses of known standard devices,
and compares the measurement with actual device data. The analyzer uses the results to
characterize the system and effectively remove the system errors from the measurement data
of a test device, using vector math capabilities internal to the network analyzer.
When you use a measurement calibration, the dynamic range and accuracy of the measurement
are limited only by system noise and stability, connector repeatability, and the accuracy to
which the characteristics of the calibration standards are known.
Application and Operation Concepts
6-57
Summary of Contents for HP 8753E
Page 61: ......
Page 150: ......
Page 187: ......
Page 197: ...4 l 8 Printing Plotting and Saving Measurement Results ...
Page 245: ......
Page 249: ......
Page 256: ......
Page 277: ...4 lb enter the number of sweeps press Optimizing Measurement Results 5 55 ...
Page 399: ......
Page 475: ......
Page 476: ...pb611ld 8 2 Menu Maps ...
Page 477: ......
Page 478: ...84 Menu Maps ...
Page 480: ......
Page 487: ......
Page 488: ...8 14 MenuMaps ...
Page 489: ...I r RETURN pg657e Menu Maps 8 15 ...
Page 565: ... Ittble 9 2 Softkey Locations Softkey Front Panel Access Key B 76 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 77 ...
Page 567: ...9 79 Key Definitions ...
Page 568: ...Key Definitions 8 78 ...
Page 569: ... Ihble 9 2 Softkey Locations continued Softkey A i _ 8 80 Key Definitions ...
Page 570: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 81 ...
Page 572: ...lhble 9 2 Softkey Lacations continued Softkey Front Panel Access Key Key Definitions 9 83 ...
Page 573: ... I ble 9 2 Softkey Locations continued Softkey Front Panel Access K e y 84 I ey Definitions ...
Page 574: ... lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions g 86 ...
Page 575: ... lhble 9 2 Softkey Locations continued Softkey FFont Panel Access Key 9 86 Kay Definitions ...
Page 576: ... ble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 87 ...
Page 577: ... lhble 9 2 Softkey Locations continued Softkey Front Fanel Access Key B 88 Key Definitions ...
Page 578: ...l hble 9 2 Softkey Locations continued Softkey Key Definitiins 8 W ...
Page 579: ... lhble 9 2 Softkey Locations continued Softkey Front Rmel Access Key WI Key Definitions ...
Page 580: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions Ml ...
Page 581: ... able 9 2 Softkey Locations continued Key Definitions ...
Page 582: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 943 ...
Page 583: ...7hble 9 2 Softkey Locations continued I Softkey Front Panel Access Key 944 Key Definitions ...
Page 584: ... Ihble 9 2 Softkey Locations continued Softkey Key Definitions 9 96 ...
Page 671: ......
Page 672: ......
Page 673: ......