is selected when the desired measurement impedance differs from the
impedance of the line standard. This requires a knowledge of the exact value of the of
the line. The system reference impedance is set using
under the calibration menu.
actual impedance of the line is set by entering the real part of the line impedance as the
. . . . .
within the
standard menu. For example, if the line was known to have
. . . . .
..... . . .
a characteristic impedance of 51
= 51 it could still be used to calibrate
50 After a calibration, all measurements would be
referenced to 50 instead of 51 When the line standard is remeasured, the center of the
Smith chart is at the current value of
(in this case, 50 Since only one value of
offset can be selected for the line standard, the value of should be a constant value over
the
frequency range of interest in order to be meaningful.
By
must have a known insertion phase or electrical length. If a non-zero length thru is specified
to have zero delay, the reference plane will be established in the middle of the thru. The
reflect standard may be used to set the reference plane instead of the thru provided the phase
response (offset delay, reactance values and standard type) of the reflect standard is known
and is specified in the calibration kit
Note
Dispersion Effects
Dispersion occurs when a transmission medium exhibits a variable propagation
or phase velocity as a function of frequency. The result of dispersion is a
non-linear phase shift versus frequency, which leads to a group delay which
is not constant. Fortunately, the TRL calibration technique accounts for
dispersive effects of the test
up to the calibration plane, provided that:
1. The thru (zero or non-zero length) is defined as having zero electrical length
. . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . .
and is used to set the reference plane
2. The transmission lines used as calibration standards have identical dispersion
characteristics (i.e., identical height, width and relative dielectric constant).
When a non-zero length thru is used to set the reference plane, it should be
as having zero length in the TRL standards
even though it
has physical length. The
electrical length of the thru standard must then
be subtracted from the actual electrical length of each line standard in the TRL
calibration kit
The device must then be mounted between two short
lengths of transmission line so that each length is exactly one-half of the length
of the non-zero length thru standard. In this
the measurement
will be properly calibrated up to the point of the device.
Application and Operation
Summary of Contents for HP 8753E
Page 61: ......
Page 150: ......
Page 187: ......
Page 197: ...4 l 8 Printing Plotting and Saving Measurement Results ...
Page 245: ......
Page 249: ......
Page 256: ......
Page 277: ...4 lb enter the number of sweeps press Optimizing Measurement Results 5 55 ...
Page 399: ......
Page 475: ......
Page 476: ...pb611ld 8 2 Menu Maps ...
Page 477: ......
Page 478: ...84 Menu Maps ...
Page 480: ......
Page 487: ......
Page 488: ...8 14 MenuMaps ...
Page 489: ...I r RETURN pg657e Menu Maps 8 15 ...
Page 565: ... Ittble 9 2 Softkey Locations Softkey Front Panel Access Key B 76 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 77 ...
Page 567: ...9 79 Key Definitions ...
Page 568: ...Key Definitions 8 78 ...
Page 569: ... Ihble 9 2 Softkey Locations continued Softkey A i _ 8 80 Key Definitions ...
Page 570: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 81 ...
Page 572: ...lhble 9 2 Softkey Lacations continued Softkey Front Panel Access Key Key Definitions 9 83 ...
Page 573: ... I ble 9 2 Softkey Locations continued Softkey Front Panel Access K e y 84 I ey Definitions ...
Page 574: ... lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions g 86 ...
Page 575: ... lhble 9 2 Softkey Locations continued Softkey FFont Panel Access Key 9 86 Kay Definitions ...
Page 576: ... ble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 87 ...
Page 577: ... lhble 9 2 Softkey Locations continued Softkey Front Fanel Access Key B 88 Key Definitions ...
Page 578: ...l hble 9 2 Softkey Locations continued Softkey Key Definitiins 8 W ...
Page 579: ... lhble 9 2 Softkey Locations continued Softkey Front Rmel Access Key WI Key Definitions ...
Page 580: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions Ml ...
Page 581: ... able 9 2 Softkey Locations continued Key Definitions ...
Page 582: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 943 ...
Page 583: ...7hble 9 2 Softkey Locations continued I Softkey Front Panel Access Key 944 Key Definitions ...
Page 584: ... Ihble 9 2 Softkey Locations continued Softkey Key Definitions 9 96 ...
Page 671: ......
Page 672: ......
Page 673: ......