Type field
The Type field selects the units in which an automatic time marker level will
be specified. The automatic time marker can have a level expressed as either
a percentage of the waveform top-base voltage (Percent) or as an absolute
voltage level (Absolute). The default selection is Percent.
at Level field
When the marker type is Percent, the at Level field sets the Tx or To marker
to a percentage level (from 10% to 90%) of the top-base voltage on the
waveform selected by the Set on field. When you select the at Level field,
you can change the percentage by turning the knob or by entering a value
using the keypad. You can enter any percentage from 10% to 90% in
increments of 1%.
When the marker type is Absolute, the at Level field sets the Tx or To marker
to the specific voltage level. The allowable voltage range that can be selected
is the vertical range for the selected channel (that is, offset
±
2 times v/div)).
You can enter any voltage from -12 V to 4 V in increments of 30 mV.
The default value for the at Level field is 50%.
The power-up default value for the at Level field is the selected channel
offset value. If the vertical range parameters (for example:v/div, offset, probe
factor) of a channel are changed such that the current at Level voltage is no
longer valid, the at Level voltage will track the limit of the vertical range.
Slope field
The Slope field sets the Tx or To marker on either the positive or negative
edge of the selected occurrence of a waveform. When you select the Slope
field, the slope toggles between Positive and Negative.
The default selection for the Slope field is Positive.
Occur field
The Occur field sets the Tx or To marker on a specific occurrence of a
displayed edge on the waveform. You can define the edge to be displayed all
the way from the 1st edge up to the 100th edge. The count of edge
occurrences is made starting with the first edge displayed on the screen.
Auto-marker measurements are made with data that is displayed on the
screen. Make sure the data of interest is fully displayed on the screen.
Any number from 1 to 100 in increments of 1 can be entered. The default
value for the Occur field is 1.
The Scope Marker Menu
Automatic time markers options
7-91
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...