Load the default calibration factors
Note that once the default calibration factors are loaded, all calibrations must
be done. This includes all of the calibrations in the Self Cal menu. The
calibration must be performed in the exact sequence listed below.
1
Go to the Scope Calibration menu.
2
Select the Mode field, then select Service Cal from the pop-up menu.
3
Select the Procedure field, then select Default Values from the
pop-up menu.
4
Select the Start field and follow the instructions on the display.
After you select the Start field, you can abort the calibration procedure by
selecting either the Mode or Procedure fields if the Continue field is still
displayed on the screen.
Calibrating the oscilloscope
Load the default calibration factors
12-8
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...