To examine software execution when a timing
violation occurs
The timing analyzer triggers when the timing violation occurs. When it
triggers, it also sets its "arm" level to true. When the state analyzer receives
the arm signal, it triggers immediately on the present state.
1
Set up one state analyzer and one timing analyzer.
2
Go to the timing analyzer’s Trigger menu.
3
Define Edge1 to represent the control line where the timing violation
occurs.
4
Under Timing Sequence Levels, enter the following sequence
specification:
•
TRIGGER on "Edge1" Occurs "1" Else on "no state" go to level "1"
5
Go to the state analyzer’s Trigger menu and check that term "a" is set
to "don’t care". In the Arming Control menu set the state analyzer to
be run by the timing analyzer.
Arming the state analyzer from the timing analyzer
6
Under State Sequence Levels, enter the following sequence
specification:
•
While storing "anystate" TRIGGER on "arm
•
a" Occurs "1" Else on "no
state" go to level "1"
Cross-Arming Trigger Examples
To examine software execution when a timing violation occurs
5-23
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...