clock derived from the sample rate oscillator. When the interpolator
indicates the stretch is complete, the counter is stopped. The count
represents, with much higher accuracy, the time between the trigger and the
first sample clock. The count is stored and used to place the recently
acquired data in relationship with previous data.
AC Cal
The AC Cal is a multiplexer circuit that can provide several
signals to the Probe Compensation/AC Calibrator output on the rear
panel. The signal provided depends on the mode of the instrument. It
can be either a probe compensation signal, a pulse representing the
trigger event, signals used for self-calibration, or the 100 MHz reference
oscillator when sample period is 1 ns.
DC Cal
The DC Cal output, a rear panel signal, is used for
self-calibration. It is one output from the 16-channel DAC.
Digital Interface
The Digital Interface provides control and interface
between the system control and digital functions in the acquisition
circuitry.
Analog Interface
The Analog Interface provides control of analog
functions in the acquisition circuitry. It is primarily a 16-channel DAC
with an accurate reference and filters on the outputs. It controls channel
offsets and trigger levels, and provides the DC Cal output.
The Analyzer Hardware
Oscilloscope board theory
9-31
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...