Minimum transitions stored
Minimum transitions stored
The figure above shows what data is stored from a data stream with
transitions that occur at a slow rate (more than 24 ns apart).
As shown, transitions are stored in two different ways, depending strictly on
chance. Remember that the transition detector only looks at the full 34 bits
while the data is stored as two 17-bit samples. So, the transition detector will
not see time tag 3 (101/000) as a transition. However, when the detector
compares the transition to time tags 2 (101/101) or 4 (000/000), it sees a
difference and detects them as transitions. For this first set of time tags, the
transition detector sees more transitions than are really there. This causes
the analyzer to store 6 samples per transition (three 34-bit sample pairs),
instead of just two, as in the 125-MHz mode. If all the transitions will be
stored in this way throughout the trace, the minimum number of stored
transitions is 682 (4096/6).
However, as you see with time tags 7 (000/000) and 8 (001/001), transitions
can fall between the pairs of samples. When this happens, only one transition
is detected and only 4 samples (two sample pairs) are stored. If all transitions
will be stored in this way, 1023 (4096/4) transitions are stored.
From run to run, the actual number of transitions stored for transitions that
occur at a slower rate will fall between these two numbers, based on the
probability of a transition falling between a sample pair or falling within a
sample pair.
Transitional Mode Theory
250-MHz transitional mode
9-9
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...