Measurement algorithms
Frequency (Freq)
The frequency of the first complete cycle displayed
is measured using the 50% levels.
If the first edge on the display is rising, then
Freq =
1
t
rising
edge
2
−
t
rising
edge
1
If the first edge on the display is falling, then
Freq
=
1
t
falling
edge
2
−
t
falling
edge
1
Period
The period is measured at the 50% voltage level of the waveform.
If the first edge on the display is rising, then
Period
=
t
rising
edge
2
−
t
rising
edge
1
If the first edge on the display is falling, then
Period
=
t
falling
edge
2
−
t
falling
edge
1
Peak-to-Peak Voltage (Vp_p)
The maximum and minimum voltages
for the selected source are measured:
Vp_p
=
V
maximum
−
V
minimum
,
where V
maximum
and V
minimum
are the maximum and minimum voltages
present on the selected source.
Positive Pulse Width (+Width)
Pulse width is measured at the 50%
voltage level.
If the first edge on the display is rising, then
+Width
=
t
falling
edge
1
−
t
rising
edg e
1
If the first edge on the display is falling, then
+Width
=
t
falling
edge
2
−
t
rising
edg e
1
Negative Pulse Width (
−
Width)
Negative pulse width is the width of
the first negative pulse on screen using the 50% levels.
If the first edge on the display is rising, then
−
Width
=
t
rising
edge
2
−
t
falling
edg e
1
If the first edge on the display is falling, then
−
Width
=
t
rising
edge
1
−
t
falling
edg e
1
The Scope Auto Measure Menu
Automatic measurement algorithms
7-100
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...