3
Optimize the Time Null of the Self Cal.
a
Connect two BNC 50-
Ω
, 9-inch cables to the BNC tee adapter.
Connect the BNC 50
Ω
(f)(f) adapter to the BNC tee adapter, and
connect the 48-inch BNC cable to the BNC 50
Ω
(f)(f) adapter. Once
you select Start, the instrument will prompt you to connect the cables
to the appropriate locations on the rear panel of the instrument.
b
Select the Procedure field, then select Time Null from the pop-up
menu.
c
Select the Start field and follow the instructions on the display.
d
After completion of the Time Null calibration, remove the cables from
the instrument.
4
Calibrate the Logic Trigger of the Self Cal.
a
Obtain a BNC 50-
Ω
, 48-inch cable.
b
Select Start. The instrument will prompt you to connect the cable to
the appropriate location on the rear panel of the instrument.
c
Select the Procedure field, then select Logic Trigger from the pop-up
menu.
d
Select the Start field and follow the instructions on the display.
e
After completion of the Logic Trigger calibration, remove the cable
from the instrument.
Calibrating the oscilloscope
Self Cal menu calibrations
12-10
Summary of Contents for 1660CS
Page 5: ...Introduction iv ...
Page 16: ...1 Logic Analyzer Overview ...
Page 24: ...2 Connecting Peripherals ...
Page 33: ...2 10 ...
Page 34: ...3 Using the Logic Analyzer ...
Page 55: ...3 22 ...
Page 56: ...4 Using the Trigger Menu ...
Page 75: ...4 20 ...
Page 76: ...5 Triggering Examples ...
Page 109: ...5 34 ...
Page 110: ...6 File Management ...
Page 119: ...6 10 ...
Page 120: ...7 Reference ...
Page 221: ...7 102 ...
Page 222: ...8 System Performance Analysis SPA Software ...
Page 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Page 252: ...9 Concepts ...
Page 284: ...10 Troubleshooting ...
Page 298: ...11 Specifications ...
Page 311: ...11 14 ...
Page 312: ...12 Operator s Service ...
Page 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Page 337: ...Glossary 4 ...