Detector System Component Description
52
Power Supply Unit Electrical Interfaces
The Synapse power supply unit
provides the following external in-
terface connections for proper system
operation:
•
AC Input Power
•
Detector Head Power
•
Power Status LED
Figure 15.
Power Supply Unit Electrical Interfaces
AC Input Power
The Synapse power supply unit operates from universal AC single-phase input power
over the range of 85 to 264 V AC with a line frequency of 47 to 63 Hz. This AC input
power is applied to a two-pole fusing power entry module located on the rear panel of the
power supply unit. This module incorporates two 5 x 20 mm IEC approved, 2.0 A, 250
V, ceramic slow blow fuses (Cooper Bussmann Part# BK / GDC-2A or equivalent) to
protect against line disturbances/anomalies outside the system’s normal operating power
range.
Detector
Head
Power
The detector head power receptacle, located on the front panel of the power supply unit,
utilizes a 16-pin circular Lemo connector to provide the required DC input power to the
detector head via the detection system’s power cable (JY# 400735).
Power
LED
Illumination of the PWR LED, located on the front panel of the power supply unit,
indicates that the unit has been powered.
Summary of Contents for JOBIN YVON SYNAPSE
Page 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Page 10: ...Preface x...
Page 14: ...System Description and Specifications 4...
Page 28: ...Detector System Installation 18...
Page 38: ...Initial Power up and Operation 28...
Page 48: ...Temperature Control 38...
Page 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Page 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Page 74: ...Appendix A Dimensional Drawings 64...
Page 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Page 100: ...Appendix E Accessories 90...
Page 104: ...Warranty 94...
Page 113: ...103 Notes...
Page 114: ...104 Notes...