Initial Power-up and Operation
27
•
ADC Speed selection
–
sets the rate at which the data is read off the CCD
detector. For maximum signal to noise, the ADC speed should be set to 20 kHz.
For maximum frame rates, the ADC speed selection should be set to 1 MHz.
•
Time Interval
–
the elapsed time between the start of one accumulation to the
start of the next accumulation. The Time Interval, Integration Time and Readout
Time of the CCD detector have the following relationship:
t
interval
≥
t
integration
+ t
read