Performing Routine Procedures
with SynerJY
CCD Focus and Alignment on the Spectrograph
1.
Attach a spectral line source, such as a mercury lamp, to the instrument entrance
slit. Consult the documentation provided with your lamp for proper mounting
instructions.
Your light source may emit high-intensity ultraviolet,
visible, or infrared light. Exposure to these types of
radiation, even reflected or diffused can result in
serious, and sometimes irreversible, eye and skin
injuries. When using a lamp, do not aim the light
guide at anyone or look directly into to the light guide
or optical ports of the instrument. Always wear
protective goggles and gloves in conjunction with the
light source.
2.
Start SynerJY. In
Experiment Setup
; select
Monos
from the
General
tab.
3.
Enter an entrance slit width of 13
μ
m (.0130 mm) and manually set the height
limiter to 1 mm.
CAUTION
Summary of Contents for JOBIN YVON SYNAPSE
Page 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Page 10: ...Preface x...
Page 14: ...System Description and Specifications 4...
Page 28: ...Detector System Installation 18...
Page 38: ...Initial Power up and Operation 28...
Page 48: ...Temperature Control 38...
Page 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Page 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Page 74: ...Appendix A Dimensional Drawings 64...
Page 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Page 100: ...Appendix E Accessories 90...
Page 104: ...Warranty 94...
Page 113: ...103 Notes...
Page 114: ...104 Notes...