37
Chapter 7: Temperature Control
Synapse monitors and regulates the array’s set point temperature via its thermostatic
control circuitry. For optimum array performance with respect to dark current, quantum
efficiency and signal-to-noise ratio, Synapse typically provides a default cooling set point
temperature -70
°
C (203 K).
Array temperature setability is provided in steps of ± 0.1
°
C resolution. From a
temperature stability standpoint, once thermal equilibrium has been reached, the
detector’s cooler power and thermostat control circuitry ensure that the array temperature
will not drift more that 0.1
°
C from the commanded value.