Initial Power-up and Operation
24
Excessive asymmetry of the peak is a sign that the slit image is not aligned to the
pixel columns; diminished shape and magnitude are symptomatic of defocusing.
7.
Stop the acquisition.
8.
Using the software, divide the chip to five equal areas.
9.
Run the experiment continuously at the initial reference wavelength.
10.
When aligned, the 5 spectra will overlap but may not show similar intensity. Each
spectrum should be 3-5 pixels wide at FWHM.
11.
To adjust the focus of the CCD camera, rotate the focus wheel with your fingers
to drive the CCD flange out from the body. To bring the camera focus in, it is
necessary to hold the camera against the wheel while rotating the focus wheel.
12.
To adjust the alignment (rotation) of the CCD camera, insert a 1.5 mm allen key
into the hole in the side of the unit to engage the CCD rotation adjustment set
screw. Turning the screw into the body (clockwise) will push against the pin on
the CCD flange rotating the camera. To rotate in the opposite direction, you need
to turn the camera against the rotation adjustment screw while turning the screw
counter-clockwise.
13.
When the focus and alignment of the camera are properly set, tighten the flange
lock to clamp the CCD flange in position.
14.
To lock the focus wheel in its current position, turn off the light source, remove
the top cover of the spectrograph and tighten the focus lock set screw. If it is
necessary to remove the CCD for some reason, simply loosen the flange lock set
screw and remove the CCD. This Quick-Align CCD adapter mechanism allows
easy replacement of the CCD with minimal realignment.
Figure 6.
Example of a Focused and Aligned CCD
Summary of Contents for JOBIN YVON SYNAPSE
Page 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Page 10: ...Preface x...
Page 14: ...System Description and Specifications 4...
Page 28: ...Detector System Installation 18...
Page 38: ...Initial Power up and Operation 28...
Page 48: ...Temperature Control 38...
Page 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Page 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Page 74: ...Appendix A Dimensional Drawings 64...
Page 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Page 100: ...Appendix E Accessories 90...
Page 104: ...Warranty 94...
Page 113: ...103 Notes...
Page 114: ...104 Notes...