Appendix C: Performing Routine Procedures with SynerJY
71
Triggering
Synapse detection systems offer both input and output TTL trigger functions. Triggering
functions are software enabled. Two hardware triggers are available as SMB receptacles
on the back of the controller: one for TTL input (male) and one for TTL output (female).
Triggering can be activated at the start of each experiment or at the start of each
acquisition during the course of one experiment.
To enable triggering using SynerJY:
1.
Start SynerJY and open the
Experiment Setup
screen.
2. Select the
Triggers
tab. The Triggers window will open.
3. From the
Input Trigger
heading, click
Enable
to activate the
Input Trigger
.
4. Select the appropriate
Input Trigger
parameters.
Event
allows the user to specify
whether the Trigger will be enabled once, at the start of the experiment or at the
Summary of Contents for JOBIN YVON SYNAPSE
Page 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Page 10: ...Preface x...
Page 14: ...System Description and Specifications 4...
Page 28: ...Detector System Installation 18...
Page 38: ...Initial Power up and Operation 28...
Page 48: ...Temperature Control 38...
Page 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Page 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Page 74: ...Appendix A Dimensional Drawings 64...
Page 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Page 100: ...Appendix E Accessories 90...
Page 104: ...Warranty 94...
Page 113: ...103 Notes...
Page 114: ...104 Notes...