2-10
2.3.3 Phase margin test
●
Perform the phase margin test of a nozzle.
●
This test is performed to check the following:
①
whether or not the nozzle is in a printable state (whether or not clogging is present) and
②
whether or not ink is deteriorated.
●
This test is performed with the IJ printer on “standby” state.
1 Press Adjustment/operational checkout on the maintenance menu.
If Adjustment/operational checkout is not displayed on the maintenance menu, set the
service personnel maintenance mode on the login user selection screen.
The adjustment/operational checkout menu is then displayed.
2 Press Phase margin test.
The phase margin test screen is then displayed.
For information on changing to the service personnel maintenance mode,
refer to “2.1 Setting of service personnel maintenance mode.”
Manual
Start
up
HOME
Back
Adjust./oper. checkout
Excitation
adjustment
Charge voltage
confirmation
Phase margin test
Heating unit
confirmation
Auto-phase gain
adjustment
Level sensor
confirmation
[Standby ]
2015.07.07 12:45
Com=0
Phase margin test
Back
[Standby ]
Adjust./oper. checkout
<Phase margin test>
Start
Proc. status: In progress
Print timing
Print data
Print phase
Button input
Sensor
For test
Data to be displayed
Automatic update
Specified phase
2015.07.07 12:45
Com=0
Start
Summary of Contents for IJ UX
Page 1: ...Service Manual HITACHI Printer Model UX Revision July 1 2015 Version ...
Page 2: ... Revision of UX service manual Revision Chapter Revised Page Frist Edition ...
Page 15: ...1 9 2 External views Rear side 1 2 2 Main body internal Appearance with inner cover T removed ...
Page 16: ...1 10 1 2 3 Print head UX E UX D B Air exhaust port ...
Page 107: ...3 27 Circuit diagram of EZJ127A ...
Page 108: ...3 28 Circuit diagram of EZJ127B ...
Page 122: ...3 42 Circuit diagram of EZJ129 ...
Page 131: ...3 51 3 3 20 Electrical Connection Diagram ...
Page 132: ...3 52 ...
Page 188: ...4 55 5 Reset the time of the R air filter to 0 on the Parts usage time management screen ...
Page 301: ...7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram UX B D 7 1 ...
Page 302: ...Circulation System Diagram UX E 7 2 ...
Page 305: ...7 5 7 2 2 Electrical Connection Diagram UX B Basic model Electrical Connection Diagram UX B ...
Page 308: ... Nozzle diameter 65μm 7 8 7 4 Dimensions around charge electrode and deflection electrode ...