11.6 Trigger Modes (Memory, Recorder, RMS Recorder Function)
―――――――――――――――――――――――――――――――――――
Trigger Screen
Waveform Screen
Flashing cursor
Selection
SINGLE
The trigger is activated once after the
START
key is
pressed. It does not repeat.
REPEAT
Trigger occurs repeatedly.
Measurement takes
place each time the trigger event occurs.
AUTO
Trigger occurs repeatedly, but if approximately
1 second elapsed without the trigger applying,
recording starts automatically. This is convenient
for checking input unit waveform (Memory recorder
function only).
11.6 Trigger Modes (Memory, Recorder,
RMS Recorder Function)
・
Except in the X-Y CONT recorder function, the trigger
mode setting can be made.
・
It determines whether or not the trigger is repeatedly
accepted and the recording is continued after the
recording operation has terminated.
Method (on the trigger screen and waveform screen)
1. Using the
CURSOR
keys, move the flashing cursor to trig
mode.
2. Select the trigger mode by using the
+
and
−
keys.