54
―――――――――――――――――――――――――――――――――――――――
5.5 Zero Adjustment
―――――――――――――――――――――――――――――――――――
Flashing cursor
NOTE
5.5 Zero Adjustment
This function provides for accurate adjustment of the
waveform to the origin position when inputting a zero
voltage. Use it for reading precise values from the screen,
a printed recording or to ensure accurate results from
waveform computations.
Method (on the status screen)
1. Using the
CURSOR
keys, move the
flashing cursor to
(0 adjust)
.
2. Press the
START
key.
・
Zero adjustment should be performed after about 30
minutes of warm-up at the powering on.
・
Zero adjustment cannot be performed while measurement
is taking place.