132
―――――――――――――――――――――――――――――――――――――――
11.3 Setting the Trigger Types
―――――――――――――――――――――――――――――――――――
Trigger Screen
Waveform Screen
×
The signal ignored
0
Low level signal
1
High level signal
Selection
Trigger Screen
Method (on the trigger screen and waveform screen)
This example illustrates settings for channel A.
The channel 2 B is set in the same manner.
Using the
CURSOR
keys, move the flashing cursor, and
make the settings.
1. Set the trigger type:
Select "
LOGIC
" by using the
+
and
−
keys.
2. Set the trigger pattern:
Sets the
logic pattern for when trigger should
take place. In case of channel A, the
settings are made for logic inputs
channel A 1 to 4 in order from the
left. Select by using the
+
and
−
keys.
3. Set the AND/OR logical operator for
the trigger pattern (trigger screen
only):
Select by using the
+
and
−
keys.
This setting cannot be set on the
waveform screen.
Summary of Contents for 8804
Page 1: ...MEMORY HiCORDER INSTRUCTION MANUAL 8804 ...
Page 2: ......
Page 28: ...6 1 2 System Operation ...
Page 40: ...18 2 1 Main Units Specifications ...
Page 102: ...80 6 8 Memory Recorder Function Possible Recording Time ...
Page 128: ...106 8 3 Basic Operational Procedures ...
Page 200: ...178 12 4 Settings on the System Screen Page 2 ...
Page 230: ...APPENDIX 8 Appendices Single Dual 2 Recorder function ...
Page 232: ...APPENDIX 10 Appendices ...
Page 237: ......
Page 238: ......