13.3 Safety Requirements
―――――――――――――――――――――――――――――――――――
8804 unit
Analog
input
H
L
GND
Maximum floating voltage:
450 V DC/AC max
8804 unit
Analog
input
H
L
GND
Voltage transformer
(PT) ground terminal
Input
PT
8804 unit
Analog
input
H
L
GND
Input
PT
When making measurements on an AC power line for
example, using a voltage transformer, ensure that the
transformer is appropriately grounded.
(a) When the voltage transformer has a ground terminal
(b) When the voltage transformer has no ground terminal
NOTE