8
―――――――――――――――――――――――――――――――――――――――
2.1 Main Units Specifications
―――――――――――――――――――――――――――――――――――
Insulation
resistance and
Dielectric strength
7 M
Ω
or more/500 VDC; one minute at
1.5 kVAC (between the main unit and the
primary side of the 9418-10)
100 M
Ω
or more/500 VDC; one minute at
2 kVAC (between the each analog input, and
between the analog input and the main unit)
Power supply
Depending on the 9418-10 AC ADAPTER (PSA-
30U-120, PHIHONG)
(powering at 12 VDC
±
10%)
Rated supply voltage of the AC adapter is
100 to 240 VAC. (Voltage fluctuations of 10%
from the rated supply voltage are taken into
account.)
Rated supply frequency is 50/60 Hz.
Six LR6 alkaline batteries are used. (The AC
adapter is prior to batteries when using
together.)
9420 BATTERY PACK can be used.
Maximum rated
power
20 VA max when using the AC adapter
15 VA max when using the batteries
Charging function
The 9420 is used.
The 9420 is in the charge mode, when the AC
adapter is connected and the power switch is
OFF.
Approx. 2 or 3 hours necessary to charge
Dimensions and
mass
Approx. 245 (W)
×
160 (H)
×
51 (D) mm
Approx. 1.2 kg