184
―――――――――――――――――――――――――――――――――――――――
13.5 Measurement Errors Caused by Signal Source Internal Resistance
―――――――――――――――――――――――――――――――――――
Rin
Rs
Es
Es: Signal voltage
Rs: Signal source
resistance
Rin: Input resistance
Measurement error
=
Es
〔
Rs
+
Rin
〕
Rin
1 -
13.5 Measurement Errors Caused by
Signal Source Internal Resistance
This problem only arises if the internal resistance of the
signal source is not sufficiently low compared to the
internal resistance of the analog input unit. If it is high,
it can lead to measurement errors.
Example:
The internal resistance of the analog input unit is
approximately 1 M
Ω
, and therefore if the signal source
resistance is 1 k
Ω
the error is increased by
approximately 0.1%.