78
―――――――――――――――――――――――――――――――――――――――
6.7 Numerical Calculation
―――――――――――――――――――――――――――――――――――
RMS value
The area value
NOTE
(5) RMS value
・
The rms value (in volts) of the waveform data is shown.
・
The equation used for the calculation is as follows:
n: number of data samples
d
i
: the ith data value of the source channel
(6) Area value
・
The value of the area from the origin position for the
waveform (the point corresponding to 0 V) to the signal
waveform is shown in volt-seconds.
・
If the A and B cursors are in use, the area between A and
B cursors is displayed.
・
The equation used for the calculation is as follows:
n : the number of data samples
d
i
: the ith data value of the source channel
h =
Δ
t: sampling period
If scaling has been set, calculation will be executed after
scaling the waveform data. (When scaling, the calculation
will be slowed down.) The unit of the waveform value is
the unit to be set. (Refer to Section 12.4.1,
Scaling
Function.")