v
―――――――――――――――――――――――――――――――――――――――
Notes on Use
―――――――――――――――――――――――――――――――――――
DANGER
To prevent the danger of electric shock or damage
to the unit, never input more than 450 V(either AC
or DC) between a pair of analog input units or
between an analog input unit and the main unit.
To prevent damage to the 8804, never exceed the
limits in the following table for the various input
connections.
Input/output terminal Maximum input voltage
Analog input
500 V DC
+
AC peak
EXT TRIG
-5 V to
+
10 V
TRIG OUT
-20 V to +30 V
500 mA max
200 mW max
When using an AC adapter, always use the
specified 9418-10 AC ADAPTER(PSA-30U-120,
PHIHONG). The rated supply voltage is 100 to 240
VAC. (Voltage fluctuations of 10% from the rated
supply voltage are taken into account.) The rated
supply frequency is 50/60 Hz. Do not use a power
supply that falls outside of this range.
Notes on Use
In order to ensure safe operation and to obtain maximum
performance from the unit, observe the cautions listed
below.