169
―――――――――――――――――――――――――――――――――――――――
12.4 Settings on the System Screen
(Page 2)
―――――――――――――――――――――――――――――――――――
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
IND
2
Y
(V)
0
X(mm)
10
1
2
B
A
NOTE
Example:
Setting the scaling for measurement
values produced when using a
displacement sensor with
characteristics as shown by the graph.
Y = 8X - 6 ... (1)
(V) (mm)
↓
X = 0.125Y + 0.75 ... (2)
(mm)
(V)
Inverting equation (1) results in equation (2). If scaling is
performed as specified, then the results of measurement,
which have been obtained as voltage values, are scaled
into displacement values in mm.
[eu/v]
[offset]
[eu]
Point A
Y = 2 V,
Displacement X
=
(+1.250E-1)
×
2 + (+7.500E-1) = 1 (mm)
Point B
Y = 10 V,
Displacement X
=
(+1.250E-1)
×
10 + (+7.500E-1) = 2 (mm)
When the numerical calculation is set, scaling will be
performed after the calculation. The rms value and area
value is, however, calculated in the value after scaling.