APPENDIX 2
―――――――――――――――――――――――――――――――――――――――
Appendices
―――――――――――――――――――――――――――――――――――
○
○
○
○
Actual input signal
Sampling period
Time
Apparent signal due to aliasing
(2) Aliasing distortion
・
If the frequency of signal being measured is significantly
higher than the sampling rate, it is possible for sampling
to produce an apparent signal which is actually non-
existent.
・
Aliasing distortion cannot be avoided with methods of
measurement, like the Memory recorder function mode, in
which the sampling period for the time axis range may
vary widely.
・
Because the measurement limit frequency (refer to next
page) is determined by the time axis range used for
measurement, try to use as high speed a range as possible
for performing measurements.
・
When measuring a repeating signal, using the auto
ranging function is another useful technique. Refer to
Section 6.4, "Auto Range Function".
Summary of Contents for 8804
Page 1: ...MEMORY HiCORDER INSTRUCTION MANUAL 8804 ...
Page 2: ......
Page 28: ...6 1 2 System Operation ...
Page 40: ...18 2 1 Main Units Specifications ...
Page 102: ...80 6 8 Memory Recorder Function Possible Recording Time ...
Page 128: ...106 8 3 Basic Operational Procedures ...
Page 200: ...178 12 4 Settings on the System Screen Page 2 ...
Page 230: ...APPENDIX 8 Appendices Single Dual 2 Recorder function ...
Page 232: ...APPENDIX 10 Appendices ...
Page 237: ......
Page 238: ......