13.6 Logic Inputs
―――――――――――――――――――――――――――――――――――
WARNING
The 8804 allows a maximum of two 9306 units to be
connected, but in this case the units have a common
ground.
NOTE
Specification
The number of Channels
4 (common ground)
Input method
Digital input
Contact input
Input resistance
50 k
Ω
or more
2 k
Ω
Threshold value
+ 1.4 V
Maximum input voltage
±
50 V
±
30 V
Response time
2
μ
s or less
NOTE
13.6.2 Logic Probe
(1) Using the 9306 LOGIC PROBE
Read the Instruction Manual for the 9306 carefully.
Voltage input and contact input are switchable. It allows
a wide range of measurements from electronic circuits to
relay timings.
The input resistance of digital input under 5 V is 100 k
Ω
.
The connector cable length is 1.5 m, and the probe tip
cable length is 20 cm.